This chapter links the basic principles of magnetic resonance with the relevant hardware components in a magnetic resonance system. These components include the magnet, radiofrequency coil(s), radiofrequency amplifier, shim coils, magnetic field gradient coils, gradient amplifiers, frequency locking circuitry, and receiver. The role of each component is explained in terms of its relationship to the associated magnetic resonance phenomenon, e.g. nuclear polarization, resonance frequency, precession, the effect of radiofrequency pulses, relaxation and spatial encoding. The requirements of each of the hardware components are quantified for experiments in three different “domains”: high-resolution liquid-state nuclear magnetic resonance spectroscopy, magic-angle spinning solid-state nuclear magnetic resonance spectroscopy, and magnetic resonance imaging. Finally, an Appendix is included, which contains details of two mathematical frameworks, namely spherical harmonics and the Biot–Savart law, used in the design of magnets, shim coils, magnetic field gradients and RF coils throughout the book.

The diversity of magnetic resonance (MR) experiments is enormous, ranging from simple one-dimensional proton nuclear magnetic resonance (NMR) spectroscopy through multi-dimensional multi-nuclear spectra to full three-dimensional magnetic resonance imaging (MRI) of morphology and function in animals and humans. Some examples of the types of data produced from different MR experiments are shown in Figure 1.1.

Figure 1.1

Examples of data produced from different magnetic resonance experiments. (a) One-dimensional proton NMR spectrum, (b) two-dimensional proton-nitrogen NMR spectrum, (c) proton image of the brain, (d) electrocardiogram-triggered proton image of the human heart, (e) phosphorus spectrum from the human calf muscle, and (f) proton image of a rat brain.

Figure 1.1

Examples of data produced from different magnetic resonance experiments. (a) One-dimensional proton NMR spectrum, (b) two-dimensional proton-nitrogen NMR spectrum, (c) proton image of the brain, (d) electrocardiogram-triggered proton image of the human heart, (e) phosphorus spectrum from the human calf muscle, and (f) proton image of a rat brain.

Close modal

Despite the widely different information content of these data, the fundamental hardware systems for NMR spectroscopy (in both the liquid and solid states) and MRI (human and animal) are very similar. The basic components include the following:

  • (i) The magnet, which polarizes the nuclei and produces a net magnetization within the sample.

  • (ii) The radiofrequency coil(s), which transmit pulses of electromagnetic (EM) energy into the sample and detect the precessing magnetization, which constitutes the MR signal.

  • (iii) Magnetic field gradient coils, which induce a spatial dependence of the nuclear precession frequency and can be used for coherence order selection, measurements of diffusion, and MRI.

  • (iv) Shim coils, which are used to produce as homogeneous a magnetic field as possible throughout the sample.

  • (v) The receiver electronics and circuitry, which amplify, filter and digitize the MR signal for data storage and post-processing.

The physical arrangement of components (i) to (iv) is shown in Figure 1.2 for a vertical bore magnet.

Figure 1.2

Schematic of the setup for high-resolution liquid-state NMR experiments. The vertical-bore superconducting magnet has a clear bore of either 89 mm (wide-bore) or 51 mm (narrow-bore). Sets of superconducting shim coils are embedded within the cryostat. Additional room temperature shims are located inside the magnet bore. The next level inwards comprises the magnetic field gradient set, which may consist of one- or three-axis gradients. The innermost structure is the RF coil containing the NMR sample.

Figure 1.2

Schematic of the setup for high-resolution liquid-state NMR experiments. The vertical-bore superconducting magnet has a clear bore of either 89 mm (wide-bore) or 51 mm (narrow-bore). Sets of superconducting shim coils are embedded within the cryostat. Additional room temperature shims are located inside the magnet bore. The next level inwards comprises the magnetic field gradient set, which may consist of one- or three-axis gradients. The innermost structure is the RF coil containing the NMR sample.

Close modal

In addition, there are a series of electronic components used to switch the gradients on and off, to produce high power RF pulses, and to amplify and digitize the signal. A simplified block diagram of a generalized MR system is shown in Figure 1.3.

Figure 1.3

Block diagram of a generic MR system. The timing of RF pulses, gradient pulses, and data acquisition are all tied to a central clock. For high-resolution NMR there are typically three or four different frequency channels (in addition to the lock channel) transmitting RF pulses, whereas for MRI usually only a proton channel is present. In high-resolution NMR a single receive coil is used: in contrast, for MRI there are usually multiple receiver coils and associated receive channels, with 32 being a typical number for commercial systems.

Figure 1.3

Block diagram of a generic MR system. The timing of RF pulses, gradient pulses, and data acquisition are all tied to a central clock. For high-resolution NMR there are typically three or four different frequency channels (in addition to the lock channel) transmitting RF pulses, whereas for MRI usually only a proton channel is present. In high-resolution NMR a single receive coil is used: in contrast, for MRI there are usually multiple receiver coils and associated receive channels, with 32 being a typical number for commercial systems.

Close modal

Table 1.1 gives an idea of the characteristics and performance of components in typical commercial NMR and MRI systems. The system performances of each of the components in the table are explained in greater detail in the relevant sections throughout the book.

Table 1.1

Characteristics of the magnet, gradients, and RF coils in commercial systems

NMRMRIMRI
  Human Animal 
Static magnetic field 4.7–23.5 tesla 1.5–9.4 tesla 4.7–21.1 tesla 
Proton frequency 200 MHz–1 GHz 63.8–400 MHz 200–900 MHz 
Gradient strength 500a mT m−1 40–80b mT m−1 1 T m−1 (6 cm i.d.) 
750 mT m−1 (9 cm i.d.) 
450 mT m−1 (12 cm i.d.) 
RF coil diameter 1.3–20 mm 10–60 cm 10–60 mm 
RF amplifier power 1 kW 30 kW 4 kW 
Shortest pulse ∼1 µs ∼10–50 µs 5–10 µs 
NMRMRIMRI
  Human Animal 
Static magnetic field 4.7–23.5 tesla 1.5–9.4 tesla 4.7–21.1 tesla 
Proton frequency 200 MHz–1 GHz 63.8–400 MHz 200–900 MHz 
Gradient strength 500a mT m−1 40–80b mT m−1 1 T m−1 (6 cm i.d.) 
750 mT m−1 (9 cm i.d.) 
450 mT m−1 (12 cm i.d.) 
RF coil diameter 1.3–20 mm 10–60 cm 10–60 mm 
RF amplifier power 1 kW 30 kW 4 kW 
Shortest pulse ∼1 µs ∼10–50 µs 5–10 µs 
a

Single axis diffusion gradients can reach ∼25 T m−1.

b

300 mT m−1 for specialized systems for the human-connectome project.28,29 

In the following sections in this chapter the basic phenomena involved in magnetic resonance are explained briefly, with links to the relevant system hardware. There are a large number of MR books dealing with the basic theory of high-resolution liquid-state NMR,1–5  solid-state NMR6–9  and MRI,10–13  and readers are advised to consult these tomes for much more in-depth analyses of different aspects of basic MR theory.

The role of the magnet is to polarize the nuclei to produce a net magnetization within the sample. For NMR spectroscopy and MRI, almost all magnets are superconducting. The magnetic field should be temporally stable and homogeneous to within parts-per-billion (ppb) throughout the sample. Most magnets are actively shielded, i.e. the fringe field does not extend significantly outside the physical dimensions of the magnet itself. Magnet design is considered in detail in Chapter 2 of this book, as well as Appendix 1A at the end of this chapter.

All nuclei with an odd atomic weight and/or an odd atomic number possess a fundamental quantum mechanical property termed “spin” and are termed “spin-active” or “NMR-active”. The most important spin-active nuclei include 1H, 13C, 15N, 23Na, 17O, 31P and 2H. Notably spin-inactive are nuclei such as 16O and 12C. Considering the proton as the simplest example, the property of spin can be viewed as the proton spinning around an internal axis of rotation giving it a certain value of angular momentum (P). Since the proton is a charged particle, this rotation results in a magnetic moment (µ). This magnetic moment produces an associated magnetic field, which has a configuration similar to that of a bar magnet. The magnitude of P is quantized in terms of the nuclear spin quantum number (I):

Equation 1.1

where h is Planck’s constant (6.63 × 10−34 Js). In the following analysis a spin 1/2 nucleus (I = 1/2) is considered, corresponding to 1H, 13C, 15N, and 31P in the previous list. In this case:

Equation 1.2

The magnitudes of the magnetic moment and the angular momentum of the proton are related by:

Equation 1.3

where γ is the nuclear gyromagnetic ratio, and has a specific value for different nuclei, with protons having the highest γ (with the exception of tritium). For protons therefore:

Equation 1.4

µ contains three components ( µx, µy and µz), each of which can have any value within the conditions governed by eqn (1.4): this situation is shown in Figure 1.4(a). However, in the presence of a strong magnetic field, B0, µz is quantized with values given by:

Equation 1.5

where mI is the nuclear magnetic quantum number, and can take values I, I − 1...−I. In the case of a proton, mI = ±1/2 and so:

Equation 1.6
Figure 1.4

(a) In the absence of a magnetic field µ lies in a random direction and the nuclei occupy a single energy level. (b) When a static magnetic field is applied, µz becomes quantized at an angle θ of 54.7° with respect to B0. Parallel (α) and anti-parallel (β) orientations correspond to two energy levels, with a difference ΔE = γhB0/2π. A greater number of nuclei occupy the lower energy, parallel state.

Figure 1.4

(a) In the absence of a magnetic field µ lies in a random direction and the nuclei occupy a single energy level. (b) When a static magnetic field is applied, µz becomes quantized at an angle θ of 54.7° with respect to B0. Parallel (α) and anti-parallel (β) orientations correspond to two energy levels, with a difference ΔE = γhB0/2π. A greater number of nuclei occupy the lower energy, parallel state.

Close modal

The orientation of µ with respect to B0 is shown in Figure 1.4(b). The interaction of the static magnetic (B0) field with µz results in Zeeman splitting, producing two energy levels: one in which µz aligns parallel to B0 (the lower energy state) and the other anti-parallel (the higher energy state).

The net magnetization, M0, of a sample containing Ns protons is proportional to the difference in populations between the two energy levels, which is dictated by Boltzmann’s equation:

Equation 1.7

Eqn (1.7) shows that the net polarization of a sample is proportional to the strength of the main magnetic field. However, since the energy difference between the two levels is very small, so is the population difference. For example, at an operating magnetic field of 11.7 tesla for every one million protons, there is a population difference of only ∼40 protons between the parallel and anti-parallel orientations.

In order to detect an MR signal, energy must be applied to the nuclear spin system to stimulate transitions between the two energy levels. A pulse of EM energy is applied at the specific resonance frequency (f0), which corresponds to the energy difference between the two levels via:

Equation 1.8

The resonance frequency in Hz, or resonance angular frequency (ω0) in radians per second, is therefore given by:

Equation 1.9

The pulsed EM wave consists of both magnetic and electric field components, and it is the magnetic component that interacts with the nuclear magnetization. The EM energy, termed a radiofrequency (RF) pulse, is transmitted via an RF coil. As shown in Figure 1.5, the magnetic field component, B1+, of the EM wave from the RF coil must be created in a direction perpendicular to B0 in order to interact with the magnetization.

Figure 1.5

(a) The effect of an RF pulse on the net magnetization is to rotate the magnetization about the axis along which B1+ is applied. (b) The angle of rotation α is given by eqn (1.10).

Figure 1.5

(a) The effect of an RF pulse on the net magnetization is to rotate the magnetization about the axis along which B1+ is applied. (b) The angle of rotation α is given by eqn (1.10).

Close modal

Using classical mechanics, the action of the RF pulse applied along one axis produces a torque perpendicular to that axis. As shown in Figure 1.5(b), the angle α by which the magnetization is rotated is proportional to the product of the strength of the applied RF field and the time, τB1, for which it is applied.

Equation 1.10

After application of an RF pulse with tip angle α about the x-axis, the net magnetization can be expressed as three different vectors.

Equation 1.11

The geometric design of the RF coil is determined by the requirement that the B1+ field produced by the coil must be perpendicular to B0. There are a large number of different geometries, which are discussed in detail in Chapter 3. One example, widely used in high-resolution liquid-state NMR, is the “saddle” coil, shown in Figure 1.6. The long axis of the coil is coincident with B0, and current flowing through the copper conductor produces a B1+ field with the required orientation. The RF coil is tuned and impedance matched to 50 Ω at the Larmor frequency for high efficiency. The RF coil may be tuned to more than one frequency, as also discussed in Chapter 3.

Figure 1.6

(a) Photograph of a saddle coil, formed of an etched copper substrate, used for high-resolution liquid-state NMR. The electrical circuitry used for tuning and impedance matching is situated below the coil. The sample is placed in the centre of the coil. (b) Current I flowing through the coil produces a B1 field that is perpendicular to B0.

Figure 1.6

(a) Photograph of a saddle coil, formed of an etched copper substrate, used for high-resolution liquid-state NMR. The electrical circuitry used for tuning and impedance matching is situated below the coil. The sample is placed in the centre of the coil. (b) Current I flowing through the coil produces a B1 field that is perpendicular to B0.

Close modal

Power to the RF coil is supplied by an RF amplifier, the general specifications of which include:

  • (i) Amplifiers for liquid-state and solid-state NMR experiments must be able to produce pulses as short as 1 µs, with accurate and reproducible shape, at frequencies up to ∼1 GHz. In solid-state NMR applications, very high power pulses with a very high duty cycle may be used for proton decoupling.

  • (ii) In the case of MRI, up to 30–60 kW of power must be available without the performance varying over time owing to any component heating, as well as the amplifiers being capable of high duty cycles for RF-intensive sequences, such as turbo spin-echoes.

The design of RF amplifiers is described in detail in Chapter 6.

When the RF pulse is turned off, the component of magnetization in the transverse plane precesses around B0, as shown in Figure 1.7(a). The precession frequency, ωprecession, is exactly the same as the frequency of irradiation:

Equation 1.12
Figure 1.7

(a) Immediately after a 90° pulse about the x-axis, the magnetization lies along the y-axis, and starts to precess at a frequency ωprecession, given by eqn (1.12). (b) In the rotating reference frame the magnetization vector appears static.

Figure 1.7

(a) Immediately after a 90° pulse about the x-axis, the magnetization lies along the y-axis, and starts to precess at a frequency ωprecession, given by eqn (1.12). (b) In the rotating reference frame the magnetization vector appears static.

Close modal

The concept of a rotating reference frame is very useful in analyzing the behaviour of the net magnetization, and is shown in Figure 1.7(b). The rotating reference frame (xy′) is defined as rotating around B0 at an angular frequency ω0.

In fact, the exact precession frequencies of different nuclei within a molecule are determined by two other factors, chemical shift and scalar coupling, which are outlined in the following sections.

The term “chemical shift” refers to the fact that protons in different chemical environments within a molecule resonate at slightly different frequencies: their precession frequencies are shifted (with respect to a reference, discussed below) with the magnitude of the shift depending on their particular chemical environment. The cause of this shift is that the exact magnetic field experienced by each proton in the molecule is slightly lower than B0 owing to the shielding effects of the electron cloud surrounding each proton. Electrons have a magnetic moment, which is opposite in sign to that of the proton, and so the effective magnetic field experienced by the proton is reduced. As an example, consider the lactic acid molecule shown in Figure 1.8. There are four different proton groups (CH3, CH, OH, COOH) in this molecule: each of these proton groups has a slightly different precession frequency since they experience a slightly different magnetic field. The effective magnetic field, Beff, experienced by a proton is given by:

Equation 1.13

where σ is called the shielding constant, and is related to the electronic environment surrounding the nucleus. The resonant frequency of the proton is given by:

Equation 1.14
Figure 1.8

(a) Chemical structure of a lactic acid molecule. (b) Ball-and-stick model of the lactic acid molecule with oxygens shown in red, protons in white and carbons in gray.

Figure 1.8

(a) Chemical structure of a lactic acid molecule. (b) Ball-and-stick model of the lactic acid molecule with oxygens shown in red, protons in white and carbons in gray.

Close modal

One of the main factors that determines the value of σ is the electronegativity of the atoms connected to the protons: the proton in an –OH group has a lower shielding constant than those in a –CH2– group since oxygen is more electronegative than carbon and pulls electrons away from the proton, thus reducing the shielding. The resonant frequency of the proton of an –OH group is therefore higher than that of the protons in a –CH2– group. The chemical shift (δ), in units of parts-per-million (ppm), is defined as:

Equation 1.15

where fref is the resonant frequency of the protons in tetramethylsilane (TMS), which acts as a reference for proton NMR spectra. Figure 1.9 shows approximate chemical shift ranges for protons in different chemical environments. For the lactic acid molecule, the order of resonance frequencies is: CH3 < CH < OH < COOH.

Figure 1.9

The range of proton chemical shifts for different chemical moieties.

Figure 1.9

The range of proton chemical shifts for different chemical moieties.

Close modal

Consider the protons in the molecule shown in Figure 1.10. There are three chemically distinct protons (HA, HB, and HC), which have three different chemical shifts and corresponding resonant frequencies. However, there is also an interaction between the two protons HA and HB, which are separated by three chemical bonds (HA-to-C-to-C-to-HB). The magnetic field experienced by HA depends upon whether HB is in the α (parallel) or β (anti-parallel) state, and vice versa: this phenomenon is termed scalar coupling. Therefore, there are four different energy levels for this coupled two-proton system: αα, αβ, βα, and ββ, as shown in Figure 1.10(b). The exact value of the scalar coupling constant J depends upon a number of different factors, including the particular type of chemical bond (single, double or triple bond), as well as the angle subtended between the C–H bonds (the Karplus angle), but in general the values are between 1 and 7 Hz. The greater the number of bonds between the protons the weaker the coupling, and so for the molecule in Figure 1.10(a) there is effectively no coupling between HA and HC since there are five chemical bonds between the protons.

Figure 1.10

(a) A chemical molecule with three different protons. HA and HB are scalar coupled, whereas HC is not coupled. (b) Energy level diagram for the coupled HA and HB protons.

Figure 1.10

(a) A chemical molecule with three different protons. HA and HB are scalar coupled, whereas HC is not coupled. (b) Energy level diagram for the coupled HA and HB protons.

Close modal

For the coupled HA–HB system, the four different transition energies corresponding to the energy level diagram in Figure 1.10(b) are given by:

Equation 1.16

Figure 1.11 shows the evolution of the precessing magnetization for the three different protons. From Figure 1.9 the chemical shift of HC has the lowest value, followed by HA and the highest value for HB. HC precesses at a single frequency, whereas protons HA and HB are “split” into two frequencies by the scalar coupling.

Figure 1.11

Evolution of the magnetization in the molecule shown in Figure 1.10 from the combined effects of chemical shift and scalar coupling. The rotating reference frame (xy′) frequency is set to the chemical shift of HA.

Figure 1.11

Evolution of the magnetization in the molecule shown in Figure 1.10 from the combined effects of chemical shift and scalar coupling. The rotating reference frame (xy′) frequency is set to the chemical shift of HA.

Close modal

The final effect that must be considered in terms of the precession of the net magnetization is relaxation. After the RF pulse has been turned off, each of the magnetization components Mz, Mx and My returns to their thermal equilibrium values, with the time-evolution determined by specific time-constants. The time-evolutions of Mz, Mx and My are characterized by differential equations known as the Bloch equations:14 

Equation 1.17

The return of Mz to its equilibrium value of M0 is governed by the spin-lattice (T1) relaxation time, and the return of Mx and My to their thermal equilibrium value of zero by the spin–spin (T2) relaxation time. It should be noted that the relative values of T1 and T2 can be very different for different types of sample, but T1 is always greater or equal to T2. For non-viscous liquids used in high-resolution NMR, the values of T1 and T2 are very similar. In contrast, in solid samples T2 can be up to six orders shorter than T1, and for human MRI the T2 of many soft tissues is between one and two orders of magnitude smaller than T1.

Solving the Bloch equations for the Mx and My components of magnetization gives:

Equation 1.18

Therefore, the precessing magnetization vectors shown in Figure 1.11 decay exponentially as a function of time. In practice, there is an addition term that gives rise to the decay of transverse magnetization, which is termed “inhomogeneous line-broadening” and has a time-constant of T2+. This term arises from spatial inhomogeneities in the magnetic field within the sample as-a-whole, and can arise from the magnet itself or more commonly with an inhomogeneous sample. The overall relaxation time for transverse magnetization is termed T2*, which is given by:

Equation 1.19

The return of Mz to its thermal equilibrium value of M0 is governed by the T1 relaxation time:

Equation 1.20

In high-resolution NMR experiments, the same RF coil used to transmit the RF pulses is also used to detect the MR signal. In contrast, in MRI experiments a large number of small coils are typically used to receive the signal, with (usually) a single large coil used to transmit the RF pulses. For simplicity, and without losing generality, Figure 1.12(a) shows the case relevant to high-resolution NMR in which the saddle coil shown in Figure 1.6 is used as the receiver coil. The time-varying magnetic field produced by the precession of the magnetization vectors results in a voltage being induced in this receiver coil. The induced voltage, V, is given by Faraday’s law and is proportional to the rate-of-change of magnetic flux:

Equation 1.21

where the coil sensitivity is defined as B1/I, i.e. the B1 field produced per unit applied current. At higher strengths of the B0 field, the protons precess at a higher frequency, the value of dϕ/dt increases, and so the detected MR signal is higher.

Figure 1.12

(a) Detection of the MR signal occurs via the voltage induced across the terminals of an RF coil by the time-varying magnetic flux produced by the precessing magnetization. (b) The induced voltage as a function of time, incorporating chemical shift and scalar coupling, as well as T2* relaxation.

Figure 1.12

(a) Detection of the MR signal occurs via the voltage induced across the terminals of an RF coil by the time-varying magnetic flux produced by the precessing magnetization. (b) The induced voltage as a function of time, incorporating chemical shift and scalar coupling, as well as T2* relaxation.

Close modal

The time-domain signal is shown in Figure 1.12(b). Based on the previous sections the signal contains a number of different frequencies owing to chemical shift and the effects of scalar coupling, and decays exponentially as a function of time. The time-domain signal is usually referred to as the free induction decay (FID).

The analogue time-domain signal shown in Figure 1.12(b) must be converted into a digital signal so that it can be stored and processed. Typically, the magnitude of the signal is on the order of millivolts or even lower, which means that it cannot be digitized directly with a high dynamic range by a commercial analogue-to-digital converter (ADC). It must first pass through the MR receiver, which is covered in detail in Chapter 7 and so only a brief overview is given here.

A block diagram of the electronic circuits in the receiver is shown in Figure 1.13. The signal first passes through a low-noise preamplifier, with a typical gain factor of 20–30 dB and noise figure ∼0.5 dB. Commercial ADCs used in MR have a maximum input voltage of 5 volts, with resolution of 16 bits, and so the voltage should be amplified to close to 5 volts before being digitized, otherwise the full dynamic range of the ADC is not utilized.

Figure 1.13

Block diagram of the individual components in an MR receiver.

Figure 1.13

Block diagram of the individual components in an MR receiver.

Close modal

High-resolution ADCs cannot sample at the very high frequencies associated with MR, and so it is necessary to demodulate the signal to reduce the sampling rate in order to take advantage of the high dynamic range. The first demodulation step uses two phase sensitive detectors (PSDs), with the inputs to the two PSDs phase shifted by 90° to produce real and imaginary outputs centred at a lower intermediate frequency, ωIF. Each PSD contains a mixer and low pass filter, with the mixer effectively acting as a signal multiplier. The output of the mixer has frequency components at (ω0 + ωIF) and (ω0ωIF): the low pass filter removes the higher frequency components. The real, sR(t), and imaginary, sI(t), components of the signal pass through a second amplification stage and are digitized using separate ADCs. In practice, digital sampling is now common in both NMR and MRI,15  and is discussed further in Chapter 7.

In NMR experiments, after the signal has been digitized and stored, it is Fourier transformed into the frequency domain (for imaging experiments, covered later in this chapter, the transformation is a multi-dimensional inverse Fourier transform). The frequency-domain signal has both real, SR(f), and imaginary, SI(f), components given by:

Equation 1.22

The real part of the frequency-domain signal, SR(f), is termed the absorption-mode spectrum, as shown in Figure 1.14. Each line in the spectrum is described by a Lorentz function with a full-width-half-maximum (FWHM) given by (πT2*)−1. The imaginary component of the frequency-domain signal is termed the dispersion-mode spectrum, as also shown in Figure 1.14. The real part of the spectrum is displayed after the phase of the spectra is adjusted to account for the finite delay between the RF pulse being switched off and the start of data acquisition. Frequency independent (zero-order) and frequency-dependent (first-order) phasing are applied so that the real part of the spectrum is completely absorptive.

Figure 1.14

Fourier transformation of the complex time-domain digitized signal gives an absorption spectrum (top) and a dispersion spectrum (below).

Figure 1.14

Fourier transformation of the complex time-domain digitized signal gives an absorption spectrum (top) and a dispersion spectrum (below).

Close modal

Despite every effort to construct a magnet that produces a completely homogeneous static field throughout the entire sample, this is impossible to achieve in practice. Finite mechanical tolerances in winding and placement of the superconducting wires, and material imperfections within the superconducting wire itself result in a small degree of spatial variation of the B0 field. Indeed, it is theoretically only possible to produce a completely homogeneous field with an infinitely long magnet, as covered in Chapter 2. Furthermore, unless the sample is a pure liquid, the different magnetic susceptibilities of components within the sample, and in particular the boundaries between these components, produce distortions in the magnetic field: this is the case particularly for MRI experiments in animals or humans, as described in Chapter 4.

In order to counteract this sample-dependent inhomogeneity in the static magnetic field, it is necessary to introduce “shim coils”, which are discussed in more detail in Chapter 4. These are “correction” coils, through which current can be passed to produce a spatially-varying compensatory magnetic field for the particular sample being studied. There are two different types of shim coils: (i) superconducting shim coils, which are situated in the cryostat of the magnet, as shown in Figure 1.2, and which are adjusted only during magnet installation, and (ii) room-temperature shims, which are normally placed on the inside of the magnet bore. In high-resolution NMR there may be up to thirty different shim coils through which current can be passed. The reason for such a high number is that it is essential to obtain narrow linewidths in high-resolution NMR spectroscopy in order to separate the resonances from protons in similar chemical environments. In MRI there are far fewer shim coils since: (i) imaging is inherently less sensitive than spectroscopy to B0 field inhomogeneities, (ii) the effect of the shims is less since they are situated a much greater distance away from the sample than in high-resolution NMR, and (iii) the intrinsic inhomogeneities produced by the sample are much greater.

The geometry of the shim coils is designed to produce different spatial variations in the magnetic field. Figure 1.15(a) shows examples of three sets of shim coils used to produce different spatial compensations to the magnetic field, Figure 1.15(b), as a function of location in the z-direction. The effects of mis-setting the shim currents are also shown in Figure 1.15.

Figure 1.15

(a) Wire geometries for Z1, Z2 and Z3 shim coils with the appropriate relative current values and directions provided by the individual shim DC power supplies. (b) The spatial variation in magnetic field in the z-direction for each of the Z1, Z2 and Z3 shim coils. (c) A spectrum showing a well-shimmed line shape. (d) A spectrum showing low-field asymmetry produced by a sub-optimal value of the Z2 shim. (e) A spectrum displaying a symmetric broad baseline produced by a sub-optimal value of the Z3 shim.

Figure 1.15

(a) Wire geometries for Z1, Z2 and Z3 shim coils with the appropriate relative current values and directions provided by the individual shim DC power supplies. (b) The spatial variation in magnetic field in the z-direction for each of the Z1, Z2 and Z3 shim coils. (c) A spectrum showing a well-shimmed line shape. (d) A spectrum showing low-field asymmetry produced by a sub-optimal value of the Z2 shim. (e) A spectrum displaying a symmetric broad baseline produced by a sub-optimal value of the Z3 shim.

Close modal

Optimization of the currents through the shim coils can be performed in two different ways. In one method, the real component of the deuterium lock signal is monitored, covered later in Section 1.9, in the other method magnetic field gradients are used, as described in Section 1.8.4.

Magnetic field gradient coils, a term usually shortened to simply “gradient coils”, are designed to produce a linear dependence of the effective magnetic field on spatial position. They consist of shaped conductors (wires or thin sheets) through which current passes to produce the gradient field. In high-resolution NMR probes, a single z-axis gradient coil is usually integrated into the RF probe, although three-axis gradients can also be present. For MRI, three independent orthogonal sets of gradient coils are present, each controlled by independent gradient amplifiers.

Since only the z-component of the magnetic field (Bz) interacts with the proton magnetic moments, it is the spatial variation of Bz that is relevant. The magnetic field gradients are designed to be linear over the sample, i.e.:

Equation 1.23

Figure 1.16 shows a plot of magnetic field vs. spatial position for a gradient applied along the z-axis.

Figure 1.16

A linear gradient in the z-direction can be produced by two sets of loops of wire (a), each one carrying current in the opposite direction. (b) Midway between the loops, at z = 0, the additional magnetic field from the two sets of loops cancels out, and the magnetic field experienced by the nuclei is B0.

Figure 1.16

A linear gradient in the z-direction can be produced by two sets of loops of wire (a), each one carrying current in the opposite direction. (b) Midway between the loops, at z = 0, the additional magnetic field from the two sets of loops cancels out, and the magnetic field experienced by the nuclei is B0.

Close modal

When the gradient is switched on by passing current through the coil, the magnetic field Bz becomes a function of z position and is given by:

Equation 1.24

where Gz has units of tesla (T) per metre. The corresponding precession frequencies (ωz) of the protons, in the rotating frame, are given by:

Equation 1.25

Details on gradient coil design, as well as the gradient amplifiers, are covered in Chapter 5. Gradient coils are used for many purposes in MR, the following being the most common:

  • (i) Dephasing or “crushing” of unwanted transverse magnetization.

  • (ii) Coherence selection in multiple quantum coherence experiments.16 

  • (iii) Measurements of apparent diffusion coefficients of liquid samples,17  or tissues in MRI.18 

  • (iv) Rapid optimization of shim currents for high-resolution NMR and MRI.

  • (v) Spatial localization in MRI19,20  and localized in vivo magnetic resonance spectroscopy (MRS).

The following sections give a brief overview of these gradient-based applications.

“Crusher” gradients are used to dephase transverse magnetization and are widely used, for example, in localized in vivo MR spectroscopy to suppress unwanted signals arising from imperfect RF pulses. In order to calculate the required strength and duration of a gradient pulse for effective dephasing, consider the effect of a gradient, Gz, applied for a time τ as shown in Figure 1.17(a), on transverse magnetization, Mxy. The resonance frequency of the protons is given by eqn (1.25). If one assumes a sample with length L in the z-direction, and with uniform proton density, then the average transverse magnetization, as a function of Gz and τ, over the length L is given by:

Equation 1.26

i.e. a sinc function. Figure 1.17(b) shows a plot of the transverse magnetization as a function of time and gradient strength. If one wishes, as an example, to completely dephase the magnetization for a sample of length 1 cm, then a 2 ms long gradient pulse with strength 0.37 T m−1 reduces the transverse magnetization by a factor of 1000.

Figure 1.17

(a) Application of a gradient pulse directly after an RF pulse dephases the magnetization. (b) The time-evolution of the transverse magnetization as a function of gradient strength.

Figure 1.17

(a) Application of a gradient pulse directly after an RF pulse dephases the magnetization. (b) The time-evolution of the transverse magnetization as a function of gradient strength.

Close modal

Many NMR experiments involve the creation and selection of multiple quantum coherences to simplify the resulting spectra. Examples include homonuclear double quantum and triple quantum coherence selection in correlated spectroscopy (COSY) experiments, as well as heteronuclear multiple quantum coherences created in triple-resonance sequences. The topic of multiple-quantum sequences is covered in many textbooks on high resolution NMR spectroscopy. Gradients can be used to select between different coherence pathways. The coherence order, p, is defined as the difference in the magnetic quantum number, m, of the relevant eigenstates, i.e.:

Equation 1.27

Longitudinal magnetization corresponds to p = 0, single quantum coherence transverse magnetization to p = ±1, double-quantum coherence to p = ±2, and so on. In any sequence, the coherence transfer pathway begins with p = 0, corresponding to thermal equilibrium magnetization along B0.

In order to select only the desired coherence pathway, phase cycling can be used, i.e. repeating the sequence a number of times with different phases of the RF pulses. This is an efficient way of coherence selection, but limits the minimum total experiment time, especially for sequences requiring extensive phase cycles (16, 32 or 64 steps). An alternative method of selecting specific coherence orders is to use gradients, using the principle that the phase accumulation of the spins during the application of the gradient pulse is proportional to the coherence order. So for a rectangular-shaped gradient pulse with strength Gz, applied for time τ, the phase ϕ is given by:

Equation 1.28

Eqn (1.28) shows, for example, that double-quantum coherences accumulate phase twice as fast as single-quantum coherence, and triple-quantum coherences three times as fast. Thus, by applying gradient pulses of different strengths or durations it is possible to refocus selective coherence orders, while unwanted coherences are dephased. Simple examples are shown in Figure 1.18 for double-quantum and triple-quantum-filtered COSY sequences.

Figure 1.18

Pulse sequences for double quantum filtered (DQF) and triple quantum filtered (TQF) magnitude COSY spectra. The sequence of RF pulses is identical for both experiments. In the DQF experiment, the double quantum coherence created by the second 90° pulse is dephased by gradient Gd and the detected single quantum coherence is rephased by gradient Gr. If Gr = 2Gd, then DQ coherence only is refocused, if Gr = 3Gd then TQ coherence only is refocused.

Figure 1.18

Pulse sequences for double quantum filtered (DQF) and triple quantum filtered (TQF) magnitude COSY spectra. The sequence of RF pulses is identical for both experiments. In the DQF experiment, the double quantum coherence created by the second 90° pulse is dephased by gradient Gd and the detected single quantum coherence is rephased by gradient Gr. If Gr = 2Gd, then DQ coherence only is refocused, if Gr = 3Gd then TQ coherence only is refocused.

Close modal

The phase evolution of transverse magnetization when a gradient is applied can also be used to measure the molecular diffusion coefficient of a sample. The most common sequence used for this purpose is the Stejskal–Tanner spin-echo, or variations thereof, shown in Figure 1.19.

Figure 1.19

The Stejskal–Tanner spin-echo experiment used to measure the apparent diffusion coefficient (D) of a sample. The value of D can be calculated from a plot of the measured signal intensities as a function of the applied gradient strength for a series of different gradient values.

Figure 1.19

The Stejskal–Tanner spin-echo experiment used to measure the apparent diffusion coefficient (D) of a sample. The value of D can be calculated from a plot of the measured signal intensities as a function of the applied gradient strength for a series of different gradient values.

Close modal

The phase accumulation during the first gradient is given by:

Equation 1.29

The 180° pulse refocuses B0 inhomogeneities and chemical shift effects. If the diffusion coefficient of the sample is zero, then there is an equal and opposite phase accumulation produced by the gradient applied after the 180° pulse, and the effects of phase accumulation are completely rephased, the signal simply being attenuated by T2 relaxation. However, if there is indeed a physical displacement of the molecules, i.e. diffusion, during the time interval between the two gradient pulses, then the two phase accumulations are not equal and the signal amplitude is attenuated. The faster the diffusion coefficient (D) the greater the signal attenuation. In the “short-pulse limit”, i.e. when δ ≪ Δ, the signal intensity is given by:

Equation 1.30

where the b-factor is defined as:

Equation 1.31

By repeating the sequence with a set of different b-values, usually via different values of the diffusion-encoding gradient, the resulting signal amplitudes can be fitted to eqn (1.30) to determine the value of D. In addition to measuring diffusion in homogeneous liquid samples, diffusion weighting can be used in high-resolution NMR21,22  to filter out signals with a very fast diffusion coefficient, for example the water signal in a protein solution. Diffusion encoding sequences are very widely used in MRI in the areas of diffusion weighted imaging (DWI),18,23  diffusion tensor imaging (DTI),24  diffusion kurtosis imaging (DKI),25,26  and fiber tractography.27 

As outlined in Section 1.7, maximizing the magnetic field homogeneity for each sample involves B0 shimming, i.e. optimizing the currents through each of the individual shim coils. This is a time-consuming process that is not ideal to perform manually, for example, for in vivo experiments. Rather than manual optimization, the process can be automated and speeded up by the use of gradients using a pulse sequence, such as the one shown in Figure 1.20(a), which measures the magnetic field inhomogeneity, ΔB0(z) as:

Equation 1.32

where Δϕ is the spatially dependent phase difference between the measured projections shown in Figure 1.20(a). In terms of the shim settings, the residual magnetic field ΔBres(z) can be defined as:

Equation 1.33

where Sj(z) is the magnetic field distribution of shim j, and cj represents the corresponding weighting function. This residual is minimized by calculating the values of the coefficients cj (i.e. the currents through each shim coil), which is normally performed in a least-squares sense over the entire one-dimensional z-projection:

Equation 1.34

where Nz is the number of data points acquired in the z-projection.

Figure 1.20

Schematic of the processes involved in gradient-based shimming. (a) One dimensional gradient projections of the sample are obtained for two different echo times, TE1 and TE2. (b) The difference in the phases (Δϕ) between the two projections is calculated and scaled to give the B0 inhomogeneity (ΔB0) as a function of z-position. (c) Knowing the magnetic field profiles of each of the shim coils, a least-square fitting routine is implemented to calculate the currents through each of the shim coils which best compensates for the measured B0 inhomogeneity.

Figure 1.20

Schematic of the processes involved in gradient-based shimming. (a) One dimensional gradient projections of the sample are obtained for two different echo times, TE1 and TE2. (b) The difference in the phases (Δϕ) between the two projections is calculated and scaled to give the B0 inhomogeneity (ΔB0) as a function of z-position. (c) Knowing the magnetic field profiles of each of the shim coils, a least-square fitting routine is implemented to calculate the currents through each of the shim coils which best compensates for the measured B0 inhomogeneity.

Close modal

In MRI three orthogonal sets of gradients are used to spatially encode the x, y and z dimensions of the sample. Each of these gradients is designed to provide a linear variation of magnetic field as a function of spatial dimension, i.e.:

Equation 1.35

The process of image formation is covered in more detail in Section 1.11 later in this chapter. For human MRI applications, typical gradient strengths of 60–70 mT m−1 are used, with some specialized gradients going up to 300 mT m−1.28,29  High gradient values can also be produced by reduced diameter “head-only insert” gradient coils. MRI is also extensively used in pre-clinical animal imaging systems, and also for microimaging experiments on samples as small as single cells.

Although the static magnetic field is extremely stable over time, every magnet displays small frequency “drifts” that are typically of the order of 1–10 Hz per hour. Since many NMR experiments require data acquisition times on the order of hours or even days, it is essential to compensate in real-time for this field drift. If not corrected, then spectral linewidths are broadened, and artifacts can appear in the spectrum.

In high-resolution liquid-state NMR, this correction is performed using a deuterium “lock channel”. One channel of one of the RF coils in the NMR probe is tuned to the deuterium (2H) resonance frequency. The sample is dissolved in a fully or partially deuterated solvent (e.g. D2O, CDCl3, or C6D6). Deuterium is a quadrupolar nucleus with a very short T1 value, and so the signal can be pulsed and sampled effectively continuously. The imaginary “dispersive” component of the deuterium spectrum is monitored, as shown in Figure 1.21(a). Changes in the static magnetic field produce a shift in the dispersive spectrum, which is analyzed to define an error signal, as shown in Figure 1.21(a). This error signal forms the input to a negative feedback loop, which controls the current applied to the Z0 shim coil. This coil is a long solenoid, which produces a uniform magnetic field which can add or subtract from the main B0 field. Since the deuterium signal can be quite noisy (owing to the low power used for pulse transmission, the low gamma of the deuterium nucleus, and the low sensitivity of the deuterium channel of the RF coil), the deuterium signal is averaged over many hundreds or thousands of samples so that the current in the Z0 shim is only updated slowly. The negative feedback loop is designed such that it integrates the deuterium signals over a long period of time, and thus has a very slow time-constant. The lock channel effectively forms a separate simple “spectrometer”, linked by a central clock/master oscillator to the rest of the data acquisition system.

Figure 1.21

Schematic of the operation of the deuterium-based frequency lock on a high-resolution NMR system. (a) A drift in the magnetic field shifts the deuterium resonance to a higher or lower frequency. The dispersive line shape is monitored and a corresponding positive or negative “error” signal recorded as the signal intensity at zero frequency in the rotating reference frame. (b) This error signal is integrated over time and, via a negative feedback loop, alters the current (ΔI) fed into the Z0 shim coil to correct for the magnetic field drift.

Figure 1.21

Schematic of the operation of the deuterium-based frequency lock on a high-resolution NMR system. (a) A drift in the magnetic field shifts the deuterium resonance to a higher or lower frequency. The dispersive line shape is monitored and a corresponding positive or negative “error” signal recorded as the signal intensity at zero frequency in the rotating reference frame. (b) This error signal is integrated over time and, via a negative feedback loop, alters the current (ΔI) fed into the Z0 shim coil to correct for the magnetic field drift.

Close modal

The lock-channel can also be used to monitor the static field homogeneity during shimming. The better the homogeneity the sharper the deuterium line shape and the higher its amplitude. This system is still present on many high-resolution NMR systems but gradient-based shimming, as described in Section 1.8.4, is increasingly being used.

For most in vivo MRI and MRS experiments there are not as stringent requirements for the stability of the magnetic field as there are in high-resolution NMR. Experiments typically do not take as long, spectral linewidths in vivo are much wider owing to the intrinsic inhomogeneity of biological tissue, and imaging per se is much more robust with respect to small frequency drifts than spectroscopy. Therefore, in MRI systems there is no need for a separate lock channel. However, in MRI there are other mechanisms that can lead to drifts in the magnetic field. In sequences requiring rapid switching of the gradients, significant temperature changes can occur in the gradient coil, which in turn can heat components such as the passive iron shims and field booster rings. These thermal disturbances change the magnetic susceptibility of the component iron, and hence cause a change in the magnetic field. Changes owing to gradient heating are relatively slow owing to the large heat capacity of the physical former on which the gradient coils are wound, and the large mass of the structure, and so monitoring does not have to be continuous but can be interspersed between scans using very low tip angle spectroscopic measurements of the water resonance frequency, since it is the dominant species for in vivo measurements. Based on these measurements, rather than changing the magnetic field itself as described for the deuterium lock, the system simply alters the frequencies of the transmitter and receiver.

Although this type of monitoring can be used to compensate for global changes in the magnetic field, it cannot correct for spatially dependent changes, which, in addition to gradient heating, can be caused by many other factors, such as patient motion. Correcting for spatially dependent changes in the magnetic field clearly requires spatially dependent measurements. These can be implemented, for example, using multiple small microcoil probes that contain a fluorinated liquid and, similarly to the deuterium lock channel on a high-resolution NMR system, form a separate but linked MR spectrometer.30  The use of multiple fluorinated probes, shown in Figure 1.22, allows the extraction of spatial variations in the main magnetic field, and the ability to apply real-time corrections to compensate for these fluctuations.

Figure 1.22

A photograph of a commercial field-monitoring system using a series of microcoils (encased in black plastic), tuned to the fluorine resonance, with a fluorinated sample inside. By continuous measurement of the phase of the fluorine MR signal, spatial and temporal variations in the magnetic field can be measured and corrected either in real-time or in signal post-processing.

Figure 1.22

A photograph of a commercial field-monitoring system using a series of microcoils (encased in black plastic), tuned to the fluorine resonance, with a fluorinated sample inside. By continuous measurement of the phase of the fluorine MR signal, spatial and temporal variations in the magnetic field can be measured and corrected either in real-time or in signal post-processing.

Close modal

NMR spectra of liquid samples consist of a series of very sharp resonances, owing to the averaging of anisotropic NMR interactions by rapid random tumbling, resulting in very long T2 values, typically hundreds of milliseconds. In contrast, NMR spectra of solid or rigid samples are very broad, since anisotropic and orientation-dependent interactions are not averaged: an example is shown in Figure 1.23(a). The two most important interactions that produce broad linewidths in solids are dipolar coupling and chemical shielding, with the dipolar term being dominant especially at low magnetic fields.

Figure 1.23

(a) Schematics of the respective spectra for 13C in the liquid and solid states. (b) Spinning the sample at the magic angle (54.7°) with respect to B0 is used to reduce the effect of dipolar coupling and chemical shift anisotropy. (c) A pulse sequence used for heteronuclear solid-state spectroscopy involving cross-polarization and high-power proton decoupling.

Figure 1.23

(a) Schematics of the respective spectra for 13C in the liquid and solid states. (b) Spinning the sample at the magic angle (54.7°) with respect to B0 is used to reduce the effect of dipolar coupling and chemical shift anisotropy. (c) A pulse sequence used for heteronuclear solid-state spectroscopy involving cross-polarization and high-power proton decoupling.

Close modal

Dipolar coupling results from the interaction of one nuclear spin with the magnetic field generated by a second nuclear spin, and vice versa. Dipolar coupling represents a direct through-space interaction, the magnitude of which is given by:

Equation 1.36

where RjkDD is the dipolar coupling between two spins j and k, and rjk is the distance between the spins.

Chemical shielding is related to the chemical shift outlined earlier for liquid samples, and is an anisotropic interaction characterized by a shielding tensor. Anisotropy in the chemical shift leads to broadening of the spectral lines. For molecules which are spatially symmetric in three-dimensions the chemical shift anisotropy (CSA) is very small. For non-symmetric molecules, the CSA is larger, and linewidths correspondingly wider.

Both dipolar coupling and chemical shielding have an angular dependence with respect to B0, containing multiplicative terms (3 cos2θ − 1), where θ is the angle between a line connecting the two nuclei and B0 for the dipolar coupling, and between the direction with the largest deshielding and B0 for the chemical shift term. One can therefore effectively average out these components by rotating the sample very rapidly at an angle of 54.74° with respect to B0 such that 3 cos2θ − 1 = 0, a technique termed magic-angle spinning (MAS),31  illustrated in Figure 1.23(b). For full averaging the rate of spinning must be greater than the anisotropic interaction. State-of-the-art microprobes are able to spin at a rate of 111 kHz.32  For regular-sized samples, typical diameter/spinning speeds are 1.3 mm diameter/65 kHz, 2.5 mm/35 kHz, 3.2 mm/23 kHz, 4 mm/8 kHz, and 7 mm/4 kHz. Since proton dipole–dipole coupling constants can be much greater than 100 kHz, full averaging is rarely achieved and MAS is often combined with multiple-pulse sequences, which also reduce the effects of dipolar coupling, based on concepts originally developed as WAHUHA33  or MREV-8.34,35  The combination of MAS with multiple-pulse sequences is called CRAMPS (Combined Rotation and Multiple-Pulse Sequence) and requires high-power amplifiers, short RF times, and fast switching between the transmit and receive mode.

For dilute spin-1/2 nuclei, such as 13C, the homonuclear (13C–13C) dipolar coupling is insignificant owing to its low occurrence. The 1H–13C heteronuclear dipolar coupling can be minimized with high-power proton decoupling,33,36  as shown in Figure 1.23(c). The sensitivity of this type of experiment can be increased using cross-polarization (CP)37  techniques, which involve polarization transfer from 1H to 13C. Efficient CP requires that the precessional frequencies in the rotating frame are the same for protons and carbon, i.e.:

Equation 1.37

This requirement implies that the B1+ fields must be spatially homogeneous at both frequencies, and as discussed in Chapter 3 this requires that the coils be very well balanced, which is challenging, particularly at very high frequencies. Cross-polarization is generally combined with relatively low-speed MAS (5 to 15 kHz) in a combined approach known as CP–MAS.

In a similar fashion to that described previously for liquid-state NMR, gradient coils can be incorporated into the probe in order to perform coherence pathway selection in multiple quantum solid-state experiments. In this case, the gradients must be oriented along the magic angle.38,39 

The instrumental requirements corresponding to the types of experiments outlined above include:

  • (i) The ability to spin the sample extremely rapidly with a high degree of mechanical precision.

  • (ii) Very short (∼1 µs), high-power RF pulses must be applied, which place high degrees of demand on the RF amplifiers and components in the RF probe.

  • (iii) The RF fields for different nuclei must be spatially very well matched and homogeneous for efficient cross-polarization.

  • (iv) Isolation between the receive chains of the different nuclei, as well as between transmit and receive chains, must be very high since high-power decoupling may be applied during data acquisition.

  • (v) Magic angle gradients and associated gradient amplifiers must be capable of producing gradient pulses with durations below 100 µs and strengths of 500 mT m−1.

In MRI there are a large number of different imaging sequences, which can be used for structural imaging, functional imaging, magnetic resonance angiography, and diffusion-weighted imaging, to name only a few.12  Two generic sequences, or “building-blocks”, are shown in Figure 1.24, namely a gradient-echo and spin-echo sequence.

Figure 1.24

(a) Generic two-dimensional gradient-echo imaging sequence. The sequence is repeated Npe times, with the phase encoding gradient being incremented from its maximum negative value to its maximum positive value. (b) Corresponding generic spin-echo imaging sequence. TR represents the repetition time and TE the echo time.

Figure 1.24

(a) Generic two-dimensional gradient-echo imaging sequence. The sequence is repeated Npe times, with the phase encoding gradient being incremented from its maximum negative value to its maximum positive value. (b) Corresponding generic spin-echo imaging sequence. TR represents the repetition time and TE the echo time.

Close modal

In the sequences shown in Figure 1.24 there are three “processes” used to produce the two-dimensional image. First, slice-selection uses a frequency-selective RF pulse applied simultaneously with a magnetic field gradient, denoted by Gslice. If the selective RF pulse, with an excitation bandwidth of ±Δωs, is applied at a frequency ωs, then protons precessing at frequencies between ωs + Δωs and ωs − Δωs are rotated from the z-axis into the transverse plane, whereas protons with precession frequencies outside this range do not experience the RF pulse, as shown in Figure 1.25(a). By changing the centre frequency (ωs) of the RF pulse, the slice can be centred at different parts of the sample. As indicated in Figure 1.25(b), the thickness of the slice can be decreased by using a larger gradient strength or a longer duration RF pulse that has a smaller frequency bandwidth.

Figure 1.25

The principle of slice selection in MRI. (a) By applying a frequency-selective RF pulse in combination with a gradient, only the nuclei within a thin “slice” experience the RF pulse and are rotated from the longitudinal axis (blue) into the transverse plane (red). (b) The thickness of the slice can be controlled by the strength of the gradient and the bandwidth of the RF pulse.

Figure 1.25

The principle of slice selection in MRI. (a) By applying a frequency-selective RF pulse in combination with a gradient, only the nuclei within a thin “slice” experience the RF pulse and are rotated from the longitudinal axis (blue) into the transverse plane (red). (b) The thickness of the slice can be controlled by the strength of the gradient and the bandwidth of the RF pulse.

Close modal

Having selected a slice, a phase encoding gradient is applied to encode an orthogonal direction via the phase of the signal. After the phase encoding gradient has been switched off, the frequency-encoding gradient (Gfreq) is turned on and data acquisition begins. Nf data points are sampled while the frequency encoding gradient is on, after which a relaxation delay (TR) occurs before the sequence is repeated using the next incremented value of the phase-encoding gradient. A total of Npe phase encoding gradients are used, resulting in a total imaging time of TR × Npe.

Assigning the Gy gradient to phase encoding, and Gx to frequency encoding, the combined effect of the phase-encoding and frequency-encoding gradients gives a signal:

Equation 1.38

where τpe is the time for which the phase encoding gradient is applied. Two variables kx and ky can be defined as:40,41 

Equation 1.39

and eqn (1.38) can be re-expressed as:

Equation 1.40

The Nf × Np data matrix can be visualized as a two-dimensional dataset as a function of kx and ky, commonly referred to as the k-space domain. Consider the Nr data points collected when the maximum negative value of the phase-encoding gradient, Gy, is applied. From eqn (1.39) the value of ky for all Nr data points corresponds to its maximum negative value. When the frequency-encoding gradient is switched on, the first data point collected corresponds to the maximum negative value of kx due to the negative dephasing lobe shown in Figure 1.24(a), the second data point to a slightly more positive value of kx and so forth, and so the Nf data points correspond to one “line” in k-space, shown as line 1 in Figure 1.26(a). The second line in k-space corresponds to the next value of the phase-encoding gradient, and so on. The spacing between the k-space points is inversely proportional to the respective spatial dimensions, or fields-of-view (FOV), in the image: Δkx = 1/FOVx, Δky = 1/FOVy.

Figure 1.26

(a) Data for MRI are acquired in “k-space”, with a Cartesian grid shown in this case. (b) The image is reconstruced by a two-dimensional inverse Fourier transform of the k-space data.

Figure 1.26

(a) Data for MRI are acquired in “k-space”, with a Cartesian grid shown in this case. (b) The image is reconstruced by a two-dimensional inverse Fourier transform of the k-space data.

Close modal

A two-dimensional inverse Fourier transform of the k-space data S(kx, ky) gives ρ(x,y), which is the MR image, as shown in Figure 1.26(b). MR images are conventionally represented as the magnitude of ρ(x,y).

Equation 1.41

The signal intensity in each voxel of the image depends on the tissue relaxation times as well as the proton density. In terms of the MR relaxation times, for an axial image acquired using slice-selection in the z-direction, the image intensity of each voxel, I(x,y), is given by:

Equation 1.42

where ρ(x,y) is the “proton density” and the tip angle of the RF pulse in the gradient echo sequence is α°. The degree of T1-weighting of the image is determined by the respective values of the tissue T1 and the chosen TR interval between repeated phase-encoding steps: if TR ≪ T1 then there is no weighting, whereas if TR ≈ T1 then the image intensity is weighted by different tissue T1 relaxation times. Similarly, the situation of TE ≈ T2* introduces T2*-weighting into the image.

Although eqn (1.42) indicates that the SNR is maximized using a value of α = 90°, this case would require a long value of TR to allow full T1 relaxation to occur, and as a result an impractically long data acquisition time for clinical applications. In order to image more rapidly, the value of α is reduced to a value considerably smaller than 90°. For a given value of TR, the value of α that maximizes the signal intensity is given by:

Equation 1.43

For example, if TR is reduced to 0.05 T1, then the optimum value of α is 8°. Using these parameters, images can be acquired in a few tens of seconds.

Gradient-echo sequences allow very rapid image acquisition. The major disadvantage is that they are weighted by tissue T2* values, which are typically much shorter than T2in vivo. In order to introduce pure T2-contrast into the image a spin-echo sequence can be used, as shown in Figure 1.24(b). The spatial encoding principles are exactly the same as for the gradient echo image, but now the intensity is weighed by the T2 rather than T2* value:

Equation 1.44

As mentioned previously there are very many types of MRI sequences. Common instrumental requirements include:

  1. RF coils and RF amplifiers must be able to handle/provide up to 30–60 kW of power, with duty cycles up to ∼20%. The pulses are both amplitude and phase modulated, with strict requirements for the reproducibility. The amplifiers must be linear over a large range of signal inputs/outputs.

  2. For rapid imaging sequences such as echo planar imaging, gradient switching rates in the hundreds of hertz to kilohertz may be necessary. In this case, one also needs to consider peripheral nerve stimulation (PNS), which can occur when the dB/dt is on the order of 50–100 T s−1.

  3. Data acquisition involves much higher bandwidths than for spectroscopic experiments: fast imaging sequences may have bandwidths up to 1 MHz for very high gradient strengths.

  4. The amount of streaming data is very high: a commercial 32-channel receive array may acquire data essentially continuously over 20 minutes, representing several gigabytes of data that has to be processed in real-time for clinical evaluation.

As seen in this first chapter, the hardware components of the MR system share many geometric properties and design goals. A cylindrical geometry is inherent in superconducting magnet design, RF volume coils, shim coils and gradient coils. Maximizing the homogeneity of the magnetic field is critical for superconducting magnet and RF coil design; producing a linear spatial variation in magnetic field is key for gradient coil design; and well-defined higher-order spatial distributions of the magnetic field are required for shim coil design. The basis for each of these component is a set of current-carrying electrical conductors. Despite the fact that the operating frequencies of these different components are quite different (DC for magnets and shim coils, tens to hundreds of kHz for gradients, and tens to hundreds of MHz for RF coils), there are common mathematical tools that can be used to produce the desired magnetic field distributions. Two of the most common formalisms, namely the Biot–Savart law and spherical harmonic decomposition, together with specific examples of their use, are described in these two appendices. These models are further refined, and more sophisticated examples given, in several of the chapters later in the book.

The fundamental bases for calculating both magnetic and electric fields are termed Maxwell’s equations, listed in eqn (1.45–1.48).

Equation 1.45
Equation 1.46
Equation 1.47
Equation 1.48

where ρv is the electric charge density, ε the permittivity, µ the permeability, H the magnetic field, E the electric field, and J the electric current density. The operators on the left of eqn (1.45) and (1.46) represent divergence, and those in eqn (1.47) and (1.48) represent curl operators. In physical terms, the divergence of a vector field describes the degree to which it behaves as either a source or a sink at a given point. If the divergence is non-zero at some point then there must be a source or sink at that position: in contrast, if the divergence is zero then it indicates that there is neither. The curl operators on the left of eqn (1.47) and (1.48) represent the spatial-variation of the electric and magnetic fields, which are coupled to the time-variation of the magnetic and electric fields on the right hand side. In other words, as an electric field propagates throughout the sample it gives rise to a time-varying magnetic field. This magnetic field, which varies as a function of space, in turn gives rise to a time-varying electric field. Eqn (1.47) and (1.48) show that magnetic and electric fields are intrinsically coupled.

Designs for magnets, gradient coils, shim coils or RF coils can be formulated by simplifying Maxwell’s equations under conditions known as “quasi-static”, which corresponds to being able to neglect any effects of the finite wavelength of the EM wave within the sample. In this case, the magnetic field can be estimated quite accurately by the Biot–Savart law, which can be derived directly from Maxwell’s equations. The Biot–Savart law describes the magnetic field (dB) at a point P distance r from the wire, produced by a current I flowing through a small section (ds) of wire. The value of dB is given by:

Equation 1.49

where µ0 is the permeability of free space, equal to 1.257 × 10−6 T m A−1. Two very simple examples are given below, a straight wire and a circular loop, both shown in Figure 1.27.

Figure 1.27

Examples of (a) a current-carrying straight wire and (b) a current-carrying circular wire-loop, for calculation of magnetic fields using the Biot–Savart law.

Figure 1.27

Examples of (a) a current-carrying straight wire and (b) a current-carrying circular wire-loop, for calculation of magnetic fields using the Biot–Savart law.

Close modal

In Figure 1.27(a), the value of dB for a straight wire can be calculated by simple trigonometry:

Equation 1.50

The total magnetic field produced by the current in the infinitely long wire is then simply given by the integral of this expression:

Equation 1.51

The result shows that the magnetic field decreases as the inverse of distance from the wire.

Consider a loop of wire with radius a, shown in Figure 1.27(b). If one considers the field produced along the central axis of the loop, there is only one non-zero component of the magnetic field, which is oriented along the z-direction and is given by:

Equation 1.52

So at the centre of the loop, the magnetic field is given by µ0I/2a, i.e. the smaller the wire loop the higher the magnetic field produced per unit current. Along the central axis of the coil, i.e. along the z-axis, the magnetic field decreases approximately exponentially with distance.

The off-axis magnetic fields can also be calculated analytically, but are more complicated. They are given (in polar coordinates) by:

Equation 1.53

where E and K are complete elliptical integrals of the second and first kinds, respectively, and k is defined as:

Equation 1.54

Using these analytical results, designs using circular loops in particular can be optimized in terms of producing the desired spatial distribution of the magnetic field. The following two examples show the application in designing a linear magnetic field gradient and a homogeneous magnetic field, both with a simple two-loop arrangement.

The simplest configuration for a coil producing a gradient in the z-direction is a “Maxwell pair”, shown in Figure 1.28(a), which consists of two separate loops of multiple turns of wire, each loop containing equal currents, I, flowing in opposite directions. In order to estimate the distance between the two loops that maximizes the linearity of the gradient, the value of Bz is first calculated using eqn (1.52):

Equation 1.55

where µ0 is the permeability of free space and a is the radius of the gradient set. By applying a Taylor series expansion, and noting that by symmetry the first two differentiable terms are zero, the first term that can give a non-linear contribution is the third derivative, given by:

Equation 1.56
Figure 1.28

Physical arrangements of wire loops for (a) a linear gradient and (b) a constant homogeneous field. The values of dgradient and dconstant are derived in terms of the coil radius a in examples A3 and A4.

Figure 1.28

Physical arrangements of wire loops for (a) a linear gradient and (b) a constant homogeneous field. The values of dgradient and dconstant are derived in terms of the coil radius a in examples A3 and A4.

Close modal

This term becomes zero at a value of d = a√3. The magnetic field produced by this gradient coil is zero at the centre of the coil, and is linearly dependent upon position in the z-direction over about one-third of the separation of the two loops. The region over which the gradient is linear can be extended by adding other sets of coils in the axial dimension, as covered later in more detail in Chapters 2 and 4.

Using the same arrangement as before, except with the currents now flowing in the same direction shown in Figure 1.28(b), the first-order term in the Taylor expansion is zero, and so the first perturbation arises from the second order term, which should therefore be set to zero, i.e.:

Equation 1.57

Combining eqn (1.55) and (1.57) gives the optimum value of the distance between the coils equal to the radius of each coil. In this case the magnetic field at the centre of the coil is given by:

Equation 1.58

As seen in Appendix A, the Biot–Savart law is an effective method for optimizing the design of simple geometries in terms of their behaviour on-axis. However, it becomes much more cumbersome to use when one needs to consider off-axis terms. In this case, a very widely used formalism is to represent magnetic fields in terms of spherical harmonics, as described in a simplified form below.

The starting point is to consider the magnetic field within a sphere of a certain diameter positioned at the centre of the magnet. If the aim is to produce a homogeneous magnetic field then one can clearly define the desired field by a single vector term Bz with fixed amplitude and direction within the sphere. As will be seen in Chapter 2, this is not possible to achieve in practice (theoretically one can achieve this only by passing current through an infinitely long solenoid) and there will be spatial perturbations in the magnetic field. As in many other areas of engineering one can define these perturbations in terms of increasing orders, i.e. first-order, second-order etc. In the case of spherical harmonics, there are three first order perturbations, five second order, seven third order, and so on. Within each order, there are a number of degrees, e.g. five degrees within the second order, seven degrees within the third order etc. The order is given the symbol n, and the degree the symbol m. The zero order term (m = 0, n = 0) represents a homogeneous magnetic field. The three first order (n = 1) perturbations correspond to linear variations in magnetic field, with the three orthogonal directions obviously being x, y and z. The degree (m) of the perturbation corresponds to the particular direction, with m = +1 being x, m = 0 being z, and m = −1 being y. The five second order terms (n = 2) correspond to m = +2, +1, 0, −1 and −2. Figure 1.29 shows the spatial distributions of the zero, first and second order spherical harmonics.

Figure 1.29

Visual representation of lower order spherical harmonics.

Figure 1.29

Visual representation of lower order spherical harmonics.

Close modal

If one considers only the terms with m = 0, one can see that these are rotationally invariant with respect to the z-axis: these are called zonal harmonics. Considering the spheres to have longitudinal and latitudinal axes, analogous to a globe, there are 2|m| zeroes in the longitudinal axis and n − |m| zeroes in the latitudinal one.

Starting from Gauss’ law of magnetism and Ampere’s law, eqn (1.47) and (1.48), one can derive a very useful equation, which is referred to as Laplace’s equation.

Equation 1.59

where the differential Laplacian operator is defined as:

Equation 1.60

In the case of an MR magnet, the magnetic field is aligned along one axis (the z-axis), and so one can simplify the equation to give:

Equation 1.61

This equation is often solved in cylindrical coordinates (the derivation is long and complicated and covered in many textbooks so is not repeated here), shown in Figure 1.30, where θ and ϕ are the polar and azimuthal angles, respectively, and r is the radius.

Figure 1.30

Cylindrical axis used for the description of spherical harmonic functions.

Figure 1.30

Cylindrical axis used for the description of spherical harmonic functions.

Close modal

The solution to Laplace’s equation results in a magnetic field B(r, θ, ϕ) which can be described by an expansion of orthogonal spherical harmonic functions Ynm and coefficients Cn,m.

Equation 1.62

where n is the order and m the degree, as outlined in the previous section. Spherical harmonic functions are defined on the surface of a sphere, and are given by:

Equation 1.63

where Cn,m and ϕn,m are constants (ϕn,m = 0 for m ≥ 0 and π/2 for m < 0), and Pnm(cos θ) are polynomial functions known as Legendre polynomial functions for m = 0, and associated Legendre polynomial functions for m ≠ 0. Legendre polynomial functions can be calculated from:

Equation 1.64

and associated Legendre polynomials via:

Equation 1.65

Combining eqn (1.62) and (1.63), the magnetic field can be expressed as:

Equation 1.66

Table 1.2 gives the mathematical values of the lower order spherical harmonics that correspond to those shown in Figure 1.29.

Table 1.2

Legendre and associated Legendre functions Pn,m(cos θ)

  P0,0(cos θ  
1 
 P1,1(cos θP1,0(cos θP1,−1(cos θ 
sinθ cosθ sinθ 
P2,2(cos θP2,1(cos θP2,0(cos θP2,−1(cos θP2,−2(cos θ
3sin2θ 3sinθcosθ ½(3cos2θ − 1) 3sinθcosθ 3sin2θ 
  P0,0(cos θ  
1 
 P1,1(cos θP1,0(cos θP1,−1(cos θ 
sinθ cosθ sinθ 
P2,2(cos θP2,1(cos θP2,0(cos θP2,−1(cos θP2,−2(cos θ
3sin2θ 3sinθcosθ ½(3cos2θ − 1) 3sinθcosθ 3sin2θ 

To relate this mathematical description to the visual one presented earlier, one can consider a number of simple cases:

  • (i) All cases with m = 0. From eqn (1.66) the magnetic field is independent of ϕ, i.e. it is rotationally invariant around the z-axis. For the particular case that n = 0, m = 0, the magnetic field is constant. When m = 0, the remaining Legendre polynomials are referred to as zonal harmonics, and give rise to what is termed a zonal magnetic field.

Equation 1.67

If θ = 0, i.e. along the z-axis:

Equation 1.68

Since Pn(1) = 1, this can be expanded to give:

Equation 1.69

This expression is very useful in determining parameters in, for example, gradient coil design as covered in the examples at the end of this Appendix.

  • (ii) When m ≠ 0, the magnetic field around the z axis “oscillates” with a frequency and initial phase ϕn,m. For the case that m = +1, the field around the circumference should exhibit one cycle, and for m = −1 also for one cycle but with a π/2 phase shift in the origin of the cycle. Similarly for m = +2 and −2, there are two cycles circumferentially, as seen in Figure 1.29. Spherical harmonics with m ≠ 0 are referred to as tesseral harmonics.

  • For n = ±1 and m = 0 the function is rotationally invariant, and there is one cycle as a function of θ.

  • For n = 1, m = ±1 the number of zeros as a function of θ is zero. These correspond to what are termed as sectoral harmonics, corresponding to the case n = |m|. Since |m| = 1 then there is one cycle around the circumference.

Having given a mathematical description of spherical harmonics, the final step is to show how these can be used in very simple examples of generating specific spatial distributions of the magnetic field. The first two cases are identical to those solved in Appendix A to show the correspondence between the two techniques.

The basic building block to generate zonal harmonics is a current loop with its central axis lying along the z-direction. A declination angle α is defined with respect to a central point z = 0 as shown in Figure 1.31.

Figure 1.31

Two wire loops producing (a) a constant homogeneous field and (b) a linear gradient. In this case, the declination angle α is solved for using a spherical harmonic approach.

Figure 1.31

Two wire loops producing (a) a constant homogeneous field and (b) a linear gradient. In this case, the declination angle α is solved for using a spherical harmonic approach.

Close modal

The magnetic field in terms of zonal harmonics is given by:

Equation 1.70

If a second loop is placed symmetrically about z = 0, then one can use the fact that:

Equation 1.71

and if equal currents are applied in the same direction in each loop, then eqn (1.71) means that all the odd zonal terms in z, z3, z5etc. are cancelled out, leaving the desired term (i.e. the homogeneous magnetic field) in z0 and the first terms that must be cancelled are in z2. From eqn (1.70) the contribution is determined by the value of P3,1(cos α). Using Table 1.2, this value is zero when 5 cos2α − 1 = 0, from which α = 63° and dconstant = a, i.e. the separation is equal to the radius, the same value (fortunately!) as derived in example A1.

Consider the same two loops of wire; if the currents in the two loops are now equal and opposite, then even orders cancel, i.e. z2 = z4 = z6 = 0. Therefore, the first term that needs to be set to zero is the third order one. From Table 1.2, P41 contains a term in (7 cos3θ − 3 cos θ), which is equal to zero for θ = 49.1°. This corresponds to a separation given by 1.73 times the radius.

As can be appreciated from Figure 1.29, generating a magnetic field gradient in either the x- or y-direction involves spherical harmonics with n = 1, m = ±1, and since m ≠ 0 these involve tesseral harmonics rather than zonal harmonics. Unlike zonal harmonics, tesseral harmonics cannot be generated by a current loop, but can be generated instead using arcs of current, as shown in Figure 1.32(a). As shown by Romeo and Hoult,42  the magnetic field produced by such an arc is given by:

Equation 1.72

where dψ represents the azimuthal width of the arc. In this case the constant terms Cn,m contain two associated Legendre polynomials. As was shown previously for rings, if two arcs are placed 180° apart with current flowing in opposite directions, then the even degrees cancel out, as shown in Figure 1.32(a). So, with respect to the degree n there are terms in B11, B13, B15etc. and with respect to the order m there are terms in B21, B31, B41etc. The term in B13 can be cancelled by appropriate choice of the width Δψ of the arc:

Equation 1.73
Figure 1.32

(a) An arc of current produces tesseral spherical harmonics. Two arcs placed 180° apart cancel out even degrees of these harmonics (note that the arcs are shown as not being physically connected). The angle 120° is chosen to cancel the B13 term. (b) By adding additional arcs symmetrically about the z = 0 plane, even orders are cancelled. (c) Physical realization of the gradient coil involves connecting the individual arcs by straight line sections, and optimizing the values of α and β for linearity.

Figure 1.32

(a) An arc of current produces tesseral spherical harmonics. Two arcs placed 180° apart cancel out even degrees of these harmonics (note that the arcs are shown as not being physically connected). The angle 120° is chosen to cancel the B13 term. (b) By adding additional arcs symmetrically about the z = 0 plane, even orders are cancelled. (c) Physical realization of the gradient coil involves connecting the individual arcs by straight line sections, and optimizing the values of α and β for linearity.

Close modal

This term is zero when the angle subtended by the arc is 120°. The next step is to cancel out the higher order terms. As shown in the previous example, by using a symmetrical arrangement about the z = 0 axis, even order terms are cancelled out by extending the number of arcs from two to four, as shown in Figure 1.32(b). The next term is the contribution from B31, which as also shown in Romeo and Hoult,42  can be made equal to zero by calculating the relevant value of Cn,m, and results in the arrangement shown in Figure 1.32(c), with the values of α and β given by 21.3° and 68.7°, respectively.

Close Modal